BENCH STUDIES OF A SMALL AXIAL PUMP FOR IMPLANTATION IN PATIENTS WITH LOW ANTHROPOMETRY

Objective: to conduct bench tests and determine the working range of the pump speed for an implanted left ventricular bypass system aimed at diagnosing and treating patients with low anthropometric status. Materials and methods. The axial pump was investigated using a custom-developed hydrodynamic test bench simulating the cardiovascular system. The bench included systems for pressure and flow measurement and registration, along with software for processing both technical and biomedical parameters. Results. The operating range of the rotor speed for the STREAM CARDIO pump required to achieve a flow rate of 2.5 ± 0.5 L/min at a pressure drop of 80 ± 5 mm Hg is 8000 ± 1000 rpm, with a power consumption of 6.5 ± 1 W. © 2025 Elsevier B.V., All rights reserved.

Авторы
Buchnev Alexander S. 1 , Kuleshov Arkadiy Pavlovich 1 , Drobyshev Alexander Alexandrovich 1 , Elenkin V.A. 1 , Shevchenko G.A. 2 , Grudinin Nikita V. 1
Издательство
Общероссийская общественная организация трансплантологов «Российское трансплантологическое общество»
Номер выпуска
2
Язык
Русский
Страницы
81-88
Статус
Опубликовано
Том
27
Год
2025
Организации
  • 1 V.I.Shumakov Federal Research Center of Transplantology and Artificial Organs, Moscow, Russian Federation
  • 2 RUDN University, Moscow, Russian Federation
Ключевые слова
axial pump; heart failure; hydrodynamic bench; left ventricular bypass; pump flow rate; pump head
Цитировать
Поделиться

Другие записи

Аватков В.А., Апанович М.Ю., Борзова А.Ю., Бордачев Т.В., Винокуров В.И., Волохов В.И., Воробьев С.В., Гуменский А.В., Иванченко В.С., Каширина Т.В., Матвеев О.В., Окунев И.Ю., Поплетеева Г.А., Сапронова М.А., Свешникова Ю.В., Фененко А.В., Феофанов К.А., Цветов П.Ю., Школярская Т.И., Штоль В.В. ...
Общество с ограниченной ответственностью Издательско-торговая корпорация "Дашков и К". 2018. 411 с.